Problem: If the device detects the pressure is higher than the set value, the device will alarm.
Possible reasons: The garment may not be applied to the body correctly or the pressure function may be faulty.
Solution: Reboot the device (Turn off, wait 30seconds and turn back on).
Disconnect the hosing from the device and wrap. Deflate the wrap by squeezing it.
Reconnect hosing attached to the device and the wrap. Ensure an audible “click” sound is heard when making the connection.
Apply the wrap to the body ensuring a form fit (not loose). Secure the wrap with Velcro strip or supplied straps.
Press START THERAPY to recommence treatment.
Press MONITOR to check pressure performance.
- If the device pressure is higher than the set value. STOP USE.
Contact Medi-Rent customer service.
- If the device continues to alarm. Turn pressure off or set to 0mmHg to continue treatment.
Contact Medi-Rent customer service.
For more detailed instructions refer to the Troubleshooting section of the owner manual.